La description

Key Facts
  • Measurement range: DC to 100 kHz
  • Resolution: 1 ?V, 100 nA, 1 m?, 1 pF, 1 Hz, 0.1 °C/F
  • Basic accuracy: 0.015 % (DC)
  • 5?-digit display (480 000 counts)
  • Simultaneous display of three measurement functions, e.g. DC + AC + statistics
  • Measurement rate: up to 200 values/s
  • Measurement functions: V (DC), I (DC), V (AC), I (AC), frequency, DC power, resistance (two and fourwire), temperature (PT100/PT500/PT1000), capacitance, diode and continuity test
  • Mathematic functions: limit testing, min./max., average, offset, DC power, dB, dBm
  • Data logging to internal memory or USB flash drive in CSV format
  • Interfaces: USB-TMC/-VCP, Ethernet, LXI, IEEE-488 (GPIB)
  • SCPI commands largely compatible with Agilent 34410A (HMC8012-G)